Specifications
Measured elements:
carbon, sulfur
Samples:
inorganic, organic
Furnace alignment:
horizontal (resistance furnace) and vertical (induction furnace)
Sample carrier:
ceramic boats / crucibles
Field of application:
agriculture, chemistry / plastics, coal / power plant, construction materials, engineering / electronics, environment / recycling, geology / mining, glass / ceramics, medicine / pharmaceuticals, steel / metallurgy
Furnaces:
induction furnace, above 2,000 °C resistance furnace (ceramic tube), adjustable up to 1,550 °C (steps of 1 °C)
Detection method:
solid state infrared absorption
Number of IR cells:
1-4
Material of IR path:
gold
Typical analysis time:
induction furnace 40-50s
resistance furnace 60-120s
Chemicals required:
copper oxide, magnesium perchlorate, sodium hydroxide
Gas required:
compressed air (4 - 6 bar / 60 - 90 psi)
oxygen 99.5 % pure (2 - 4 bar / 30 - 60 psi)
Power:
Induction Furnace: 230 V, 50/60 Hz max. 15 A, 3450 W
Resistance Furnace: 230 V, 50/60 Hz, max. heating up current 20 A
Dimensions (WxHxD):
88 x 80 x 60 cm
Weight:
~ 150 kg
Required equipment:
balance (resolution 0.0001g), monitor, PC
Optional accessories:
Autoloader (for 36 or 130 crucibles), carrier gas purification, halogen trap, HTF-540 pre-heating furnace, voltage stabilizer 5 KVA